Alcatel Adixen Turbo Pumps

The Alcatel turbo pumps consist of a comprehensive line of units with the MDP and ATP designs. Made exclusively from carefully selected machined aluminum with an uncompromising architecture, and low rotational speed, they are designed to resist difficult applications. With pumping speeds from 7.5 to 900 liters a second they are often found in industrial applications, instruments, and R&D experiments where reliability, performance and cost effectiveness are essential. The ATP series turbo pump design is focused on reliability, robustness and minimum downtime for maintenance.

Features
• Ceramic Bearings
• Air/Water cooling
• Mount in any Orientation
• ISO, CF, ASA flange choice
• Option of Corrosive Version

Technical Specifications
  MDP5011 ATP80/80C ATP100/100C ATP150/150C ATP400/400C ATP900/900C
Inlet Flange
 
IS0 63
ISO 63
CF 63
ISO 100
CF 100
ISO 100
CF 100
ISO 160
CF 160
ISO 200
CF 200
Pumping Speed, L/S
Nitrogen
Helium
Hydrogen
7.5
4
3
80
50
40
100
60
40
140
100
80
400
300
250
900
540
300
Compression Ratio
 
10/9
10/4
10/3
8 x 10/7
2500
300
8 x 10/7
2500
300
7 x 10/8
1.2 x 10/4
1 x 10/3
7 x 10/8
1.5 x 10/4
1 x 10/3
7 x 10/8
1.5 x 10/4
1 x 10/3
Ultimate Pressure (mbar)
 
10 - 6
10 - 9
10 - 9
10 - 10
10 - 10
10 - 10
Startup Time (min)
 
1
2
2
2
3
3
Rotational Speed (rpm)
 
27,000
27,000
27,000
27,000
27,000
27,000
Orientation
 
Any
Any
Any
Any
Any
Any
Backing Pressure (mbar)
 
20
2x10-1
3x10-1
4x10-1
2x10-1
3x10-1
Outlet Flange
 
NW16
NW25
NW25
NW25
NW40
NW40
Weight (lb)
 
7
10
12
14
20
40
Price
 
$3,020.00
$3,830.00
$5,000.00
$6,370.00
$12,000.00
$16,390.00
Controller
 
ACT100
ACT200TH
ACT200TH
ACT600TH
ACT600TH
ACT600H
Price


$1,480.00
$1,750.00
$1,750.00
$2,720.00
$2,720.00
$2,720.00

ACT Controllers

The ACT controllers offer convenient and comprehensive electronic monitoring of your turbo pump. The modern pumping monitor displays rotational speed, pump current consumption, pump and controller temperature, total running time, pump fault detection, and diagnostic mode with last ten alarm codes. Each unit has RS 232/485 configuration. Outputs include; pump starting, pump at speed, stand-by, venting valve on/off, pump supply to air cooling fan, and selectable 0-10volts output.




Ordering Information
  MDP5011 ATP80/80C ATP100/100C ATP150/150C ATP400/400C ATP900/900C
Turbo Pump Systems ISO
Air Cooling
Water Cooling
795601
795604
F12101
F12121
G13111
G13121
H13111
H13121
I14111
I14121
K15111
K15121
Turbo Pump Systems CF
Air Cooling
Water Cooling
F12201
F12211
G13211
G13221
H13211
H13221
I14211
I14221
K15211
K15221
Controller
 
111156
20004
20004
11692
11692
11692
Pump Cable (3.5m)
 
A461868-035
A463207-035
A463207-035
A461237-035
A461237-035
A461237-035
Power Cord
 
14487
Inlet Screen
ISO
CF
063117
063000
063115
056844
056845
056844
056845
056942
056928
063158
063159
Vent Vale 115V
 
063171
063089
063089
063089
063099
063099